High Temperature Extraction Method CEMS Monitoring System For SO2 NO NO2 Fast

  1. MOQ: 1 Unit
  2. Price: USD10000-20000
  3. Get Latest Price
SO2 Range 0-100mg/m3 OEM
Response Time ≤120S
Linearity Error ≤±2%F.S.
O2 Range 0-25%
Payment Terms T/T,Online order
Repeatability ≤2%
Supply Ability 100 Unit/Week
Delivery Time 10-15 working days
NO Range 0-100mg/m3 OEM
Packaging Details Neutral carton
NO2 Range 0-100mg/m3 OEM
Product name CEMS for SO2 NO NO2
Model Number AG-CEMS09
Place of Origin Nanjing, China
Brand Name JUGE
Certification IOS9001, CE

View Detail Information

Inquiry by Email Get Latest Price
Chat online Now Ask for best deal
  1. Product Details
  2. Company Details

Product Specification

SO2 Range 0-100mg/m3 OEM Response Time ≤120S
Linearity Error ≤±2%F.S. O2 Range 0-25%
Payment Terms T/T,Online order Repeatability ≤2%
Supply Ability 100 Unit/Week Delivery Time 10-15 working days
NO Range 0-100mg/m3 OEM Packaging Details Neutral carton
NO2 Range 0-100mg/m3 OEM Product name CEMS for SO2 NO NO2
Model Number AG-CEMS09 Place of Origin Nanjing, China
Brand Name JUGE Certification IOS9001, CE
High Light OEM continuous stack emission monitoring systemNO2 continuous stack emission monitoring systemSO2 cems solutions
High Temperature Extraction Method CEMS Monitoring System For SO2 NO NO2 Fast Response Gas Analyzer
System Overview

This Continuous Emission Monitoring System (CEMS) monitors real-time concentrations of particulate matter and gaseous pollutants in exhaust gases from fixed pollution sources including boilers, industrial furnaces, and incinerators. The system is specifically designed for high-humidity, low-concentration environments.

The CEMS consists of three main components: the gaseous pollutant monitoring unit (SO2, NO, NO2), the flue gas parameter monitoring unit (temperature, pressure, flow rate, O2, humidity), and the data acquisition and processing unit.

Sample gas passes through a high-temperature heating pipeline, is filtered by a two-stage dust filter, and directly enters the gas chamber in the analyzer. The gas chamber is placed inside a high-temperature box and connected via UV fiber optics to a UV differential analyzer for precise flue gas measurement.

Key Features
  • Full-flow high-temperature extraction design with fast response time, accurate measurement, and minimal sulfur loss
  • PLC control with adjustable parameters for complex operating conditions
  • Fully intelligent system design with fault alarms and easy maintenance
  • Fully modular design allowing flexible expansion of monitoring factors
  • Direct NO measurement, avoiding conversion loss and ensuring accurate results
Technical Parameters
Gaseous Pollutants
SO2 (0~200) mg/m³, customizable range
NO (0~200) mg/m³, customizable range
NO2 (0~200) mg/m³, customizable range
O2 (0~25)%
Repetitiveness ≤2%
24h zero drift and range drift ≤±2% f.s.
One-week zero-point drift and range drift ≤±3% f.s.
Linear error ≤±2% f.s.
Response time ≤120s
Flue Gas Parameters
Temperature Measuring range (0~300) ºC, customizable
Accuracy ≤±3℃
Pressure Measuring range (-10~10) kPa
Accuracy ≤0.5%
Velocity of flow Measuring range (0~40) m/s, (0~2) kPa, output optional
Accuracy ≤±2%
Humidity Measuring range (0~40) vol%
Accuracy ≤±2%
O2 Measuring range (0~25)%
Accuracy ±2%
Data Acquisition and Processing
Data transmission module 8-way analog input, 1-channel RS485 communication interface
IPC 6-way RS232 communication interface, Windows operating system
Operational software Flue gas emission monitoring software monitor
Power Supply AC 220V, 50Hz, ≤7kW
System Composition

Company Details

Bronze Gleitlager

,

Bronze Sleeve Bushings

 and 

Graphite Plugged Bushings

 from Quality China Factory
  • Business Type:

    Manufacturer

  • Year Established:

    2015

  • Total Annual:

    2000000-3000000

  • Employee Number:

    50~100

  • Ecer Certification:

    Verified Supplier

Reliable Solutions · Flexible Cooperation · Global Delivery · Project-Oriented Support · Long-Term Partnership Nanjing Juge Environmental Technology Co., Ltd. is a global provider of industrial environmental monitoring solutions, specializing in Continuous Emission ... Reliable Solutions · Flexible Cooperation · Global Delivery · Project-Oriented Support · Long-Term Partnership Nanjing Juge Environmental Technology Co., Ltd. is a global provider of industrial environmental monitoring solutions, specializing in Continuous Emission ...

+ Read More

Get in touch with us

  • Reach Us
  • Nanjing Juge Environmental Technology Co., Ltd.
  • Building 17, No. 99 Guangli Road, Moling Subdistrict, Jiangning District, Nanjing, Jiangsu
  • https://www.cems-system.com/

Leave a Message, we will call you back quickly!

Email

Check your email

Phone Number

Check your phone number

Requirement Details

Your message must be between 20-3,000 characters!

Submit Requirement