| Payment Terms | L/C, T/T, Western Union |
| Supply Ability | 100 sets per month |
| Delivery Time | 10 working days |
| Packaging Details | carton, wooden case |
| Display mode | LCD display |
| Product Name | Large screen patient monitor |
| Battery type | Lithium battery 11.1V/4.4Ah |
| Recharge time | ≥4 hour |
| Use time | 3 hour |
| Environment temperature | 5℃~+40℃ |
| Brand Name | STATEMED |
| Model Number | STT15-MPM44 |
| Certification | CE |
| Place of Origin | CHINA |
View Detail Information
Explore similar products
Automated Portable Multi Parameter Patient Monitor / All In One Vital Sign
Small Volume Electronic Vital Sign Machine , CHD Room Vital Sign Monitoring
7 Inch Mobile Multi Parameter Patient Monitor With High Definition LCD Screen
Blood Pressure Multi Parameter Patient Monitor Light Weight Medical Equipment
Product Specification
| Payment Terms | L/C, T/T, Western Union | Supply Ability | 100 sets per month |
| Delivery Time | 10 working days | Packaging Details | carton, wooden case |
| Display mode | LCD display | Product Name | Large screen patient monitor |
| Battery type | Lithium battery 11.1V/4.4Ah | Recharge time | ≥4 hour |
| Use time | 3 hour | Environment temperature | 5℃~+40℃ |
| Brand Name | STATEMED | Model Number | STT15-MPM44 |
| Certification | CE | Place of Origin | CHINA |
| High Light | vital signs monitoring devices ,patient monitoring equipment | ||
Doctor Diagnose Portable Multi Parameter Patient Monitor CE Approved
Battery
| Battery type | Lithium battery 11.1V/4.4Ah |
Recharge time | ≥4 hour |
| Use time | 3hours |
Work Environment
| Environment temperature | Working 5℃~+40℃ |
| Relative humidity | Working ≤85% |
| Atmospheric pressure | Working 700hpa~1060hpa |
ECG
| Lead | Standard Three leads | R,L,F, lead mode :Ⅰ,Ⅱ,Ⅲ | |||
| Standard Five leads | R,L,F,N,C, | ||||
| Sensitivity | 2.5mm/mV,5mm/mV,10mm/mV,20mm/mV,40mm/mV,erro≤±3% | ||||
| Scan speed | 12.5,25,50mm/s three level,erro ≤±10% | ||||
| Adjustable screen aspect ratio | 0.05~3.2 | ||||
| Lead off condition detected | U=2V I=0.04uA | ||||
| Active noise suppression | In patients who have not applied current | ||||
| Tall T-wave suppression | 1.4mV | ||||
| Inner voice | <30μVP-P | ||||
| Input impedance | ≥2.5MΩ | ||||
| Input loop current | ≤0.1μA | ||||
| Endure electrode polarizing voltage | ±300mV | ||||
| ECG signal input range | ±5mV | ||||
| Measurement mode | Diagnosis mode,monitoring mode,operation mode | ||||
| Frequency property | diagnosis mode 0.05~130HZ | ||||
| CMRR | Diagnosis mode ≥80dB | ||||
| HR measurement range | Adult 30~250bpm | ||||
| HR accuracy | ±2% | ||||
| HR sensitivity | >200μV | ||||
| Heart rate calculation | Take 4 to 12 R-wave cycle averaged | ||||
| Refresh time heart rate | 1S | ||||
| Heart rate response time | 80-120 BPM:range 4-7s | ||||
| Alarm range | 0~301bpm,upper and lower limit can be adjusted continuously | ||||
| Alarm indicator | Audial and visual alarm | ||||
| Heart rate too fast startup time of the alarm | Range :4-6s | ||||
| Protection | Isolated immunity voltage 4000V,50/60HZImmunity of electronic surgical Anti-defibrillation | ||||
| Protection and recovery for defibrillation effect | In the monitor and the surgical model ,ECG waveforms recovery to baseline in 5S. | ||||
| Electrode polarization recovery time after the fibrillation | In the monitor and the surgical mode ,ECG waveforms recovered to baseline in 10S. | ||||
| Surgical interference of electrical recovery time | In the monitor and the surgical mode ,ECG waveforms recovered to baseline in 10S. | ||||
| No overshoot pacemaker pulse suppression: | |||||
| Extent±2mV~±700mV,Pulse width 0.1ms~2ms | QRS wave | Display | |||
| a)Single pacing pulse | 60 | 60±2bpm | |||
| 30 | 30±2bpm | ||||
| 0 | -- | ||||
| b)With normal QRS-T pacing the pacing pulse | 60 | 60±2bpm | |||
| 30 | 30±2bpm | ||||
| 0 | -- | ||||
| c)QRS has an invalid pacing mode pacing pulse. In all cases, overshoot (a0) pacing rate should be less than 5% overshoot and settling time must be less than 5μs. Rise and fall time of the pulse width should be 10% but not more than 100μs, and the pacing pulse starting time should be the starting point in the QRS complex prior to or shorter than 40ms at 40ms. | 60 | 60±2bpm | |||
| 30 | 30±2bpm | ||||
| 0 | -- | ||||
| Beware of the room pacing pulses have the same ventricular pacing pulse amplitude and duration, and ventricular pacing pulses than 150ms to 250ms in advance | 60 | 60±2bpm | |||
| 30 | 30±2bpm | ||||
| 0 | -- | ||||
| Typical minimum input slew rate | 50V/s RTI | ||||
![]()
Company Details
Business Type:
Manufacturer,Agent,Exporter,Trading Company,Seller
Year Established:
2002
Total Annual:
1-1.5 million
Employee Number:
1~200
Ecer Certification:
Active Member
State (Beijing) Medical Technology Co., Ltd. established in 2002, is a High-tech medical products enterprise combining R&D, production ,sales and after-sale together, specialized in Medical Ultrasonic Instruments, ENT Workstation, Medical Endoscope Imaging Systems, Colposcope, Spectral Instrumen... State (Beijing) Medical Technology Co., Ltd. established in 2002, is a High-tech medical products enterprise combining R&D, production ,sales and after-sale together, specialized in Medical Ultrasonic Instruments, ENT Workstation, Medical Endoscope Imaging Systems, Colposcope, Spectral Instrumen...
Get in touch with us
Leave a Message, we will call you back quickly!