Packaging Details | Neutral carton |
Delivery Time | 5-8 working days |
Payment Terms | Online order,T/T |
Supply Ability | 100 Unit/Week |
Type | Iron |
Principle | Phenanthroline spectrophotometric method |
Range | 0-20mg/L OEM |
Accuracy | ≤±10% |
Repeatability | ≤±5% |
Measurement Period | 30min |
Output | RS-232, 4-20mA, RS-485 |
Specification | 1500*500*450mm |
Certification | IOS9001, CE |
Brand Name | JUGE |
Model Number | AG-TFe07 |
Place of Origin | Nanjing, China |
View Detail Information
Explore similar products
Online SDI Analyzer SDI15 Sludge Density Index Analyzer For RO System
Nessler Spectrophotometric Method Ammonia Nitrogen Water Quality Online
Online CODcr COD Water Quality Analyzer Potassium Dichromate Oxidation
Industrial Sewage Online Ammonia nitrogen Water Quality Analyzer
Product Specification
Packaging Details | Neutral carton | Delivery Time | 5-8 working days |
Payment Terms | Online order,T/T | Supply Ability | 100 Unit/Week |
Type | Iron | Principle | Phenanthroline spectrophotometric method |
Range | 0-20mg/L OEM | Accuracy | ≤±10% |
Repeatability | ≤±5% | Measurement Period | 30min |
Output | RS-232, 4-20mA, RS-485 | Specification | 1500*500*450mm |
Certification | IOS9001, CE | Brand Name | JUGE |
Model Number | AG-TFe07 | Place of Origin | Nanjing, China |
High Light | Pollutant Source Monitoring Water Quality Analyzer ,Pollutant Source Monitoring online water analyzer ,Total Ferric Iron in Waste Water Analyzer |
This method is applicable to wastewater with Total iron concentrations within the range of 0-20 mg/L.
1. Measurement Method: Phenanthroline spectrophotometric method
2. Measurement Range: 0-5 mg/L, OEM 0-20mg/L
3. Accuracy: ≤±10%
4. Repeatability: ≤±5%
5. Measurement Cycle: Minimum cycle is 30 minutes; color development time can be adjusted flexibly between 5-30 min based on environmental conditions.
6. Sampling Cycle: Adjustable time interval (10-9999 min) or on-the-hour measurement mode.
7. Calibration Cycle: Adjustable at any interval from 1-99 days.
8. Maintenance Cycle: ≥168 hours, with each maintenance lasting ~1 hour.
9. Reagent Consumption: <0.50 CNY per sample.
10. Outputs: RS-232, 4mA-20mA, RS-485.
11. Environmental Requirements: Temperature-controlled indoor environment (recommended: 5-40°C); humidity 45%-85%.
12. Power Supply: AC 220±10% V, 50±1% Hz, 5A.
13. Dimensions: Height 1500 × Width 500 × Depth 450 mm.
14. Additional Features:
Unique Design: This product features a distinctive design that ensures lower failure rates, reduced maintenance requirements, minimal reagent consumption, and higher cost-effectiveness compared to similar products.
Selector Valve Assembly:
Micro-Dosing Unit:
Sample Introduction Unit:
Sealed Digestion Unit:
Reagent Tubing:
The ferric iron (Fe³) in the water sample is reduced to ferrous ions (Fe²) by hydroxylamine hydrochloride. Under the action of a buffer solution, the ferrous ions react with 1,10-phenanthroline to form a colored complex. The analyzer detects this color change, converts the variation into a total iron concentration value, and outputs the result. The amount of colored complex formed corresponds to the total iron content in the sample.
Company Details
Business Type:
Manufacturer
Year Established:
2015
Employee Number:
50~100
Ecer Certification:
Verified Supplier
Founded in 2015, Juge focuses on the field of environmental online monitoring and has gathered outstanding talents who have been engaged in environmental online monitoring for many years. At the same time, it cooperates with some key professional universities and research institutes in China, and is... Founded in 2015, Juge focuses on the field of environmental online monitoring and has gathered outstanding talents who have been engaged in environmental online monitoring for many years. At the same time, it cooperates with some key professional universities and research institutes in China, and is...
Get in touch with us
Leave a Message, we will call you back quickly!