{"title":"Wenner Sensor Hardened Concrete Resistivity Tester with ASTM C1760","imgUrl":"https:\/\/img.chinax.com\/nimg\/80\/11\/18687c35cd22472f2e56087f60b5-200x200-1\/wenner_sensor_hardened_concrete_resistivity_tester_with_astm_c1760.jpg","attrs":{"Brand Name":"GOLD","Certification":"ISO","Place of Origin":"China","Minimum Order Quantity":"1 set"}}
{"title":"High precision ASTM C1202 Concrete Resist Chloride Ion Penetration Tester","imgUrl":"https:\/\/img.chinax.com\/nimg\/9b\/24\/a32692a115697f78786874f68607-200x200-1\/high_precision_astm_c1202_concrete_resist_chloride_ion_penetration_tester.jpg","attrs":{"Brand Name":"GOLD","Model Number":"GD-ASTM C1202","Certification":"ISO","Place of Origin":"CHINA"}}
{"title":"NT BUILD 492 Concrete Chloride Ion Diffusion Coefficient Tester","imgUrl":"https:\/\/img.chinax.com\/nimg\/c3\/57\/79a5c63ec19956e07a7f2a0dee71-200x200-1\/nt_build_492_concrete_chloride_ion_diffusion_coefficient_tester.jpg","attrs":{"Brand Name":"GOLD","Model Number":"GD-NT BUILD 492","Certification":"ISO","Place of Origin":"CHINA"}}
{"title":"ASTM C1760 Hardened Concrete Resistivity Tester for Bulk Electrical Conductivity","imgUrl":"https:\/\/img.chinax.com\/nimg\/72\/1d\/d101453970ab6a3d01b39336c9a3-200x200-1\/astm_c1760_hardened_concrete_resistivity_tester_for_bulk_electrical_conductivity.jpg","attrs":{"Brand Name":"GOLD","Model Number":"GD-ASTM C1760","Certification":"ISO","Place of Origin":"CHINA"}}
ASTM C1202 Rcpt Concrete Chloride Penetration Teste
1.Application:
Concrete reinforcement corrosion is the primary factor affecting the durability of concrete. According to statistics, the durability damage of reinforced concrete caused by corrosion accounted for more than 40% of all injuries, while the chloride ion erosion is a major incentive for reinforcement corrosion. DTL concrete chloride ion electric flux detector can be an important indicator for the durability of concrete - concrete to resist chloride ion penetration properties (density) measurements, test indicator is the six-hour power flux (Coulomb)for the concrete specimens, quickly to evaluate the concrete resistance to the chloride ion penetration ability
2.Standard:
GB/T 50082-2024 Standard for test methods of long-term performance and durability of ordinary concrete
The Provisional Technical conditions of Passenger special line high-performance concrete
JTJ275-2000 Corrosion prevention technical specification for concrete structure of marine harbour engineering
JG/T261-2009 Apparatus to determine chloride coulomb electric flus of concrete
ASTM C1202-12 Standard Test Method for Electrical Indication of Concrete's Ability to Resist Chloride Ion Penetration
3.Detection method:
electricity
4.Product features:
Embedded computer control, power outages and power-down data storage, advanced technology, well-made, reliable performance, high accuracy
LCD display, with real-time display/print and analysis function, can be separated from Pc to use directly, to facilitate the outdoor field detection.
Standard RS232 serial communications with fully automatic PC connection software can output the flux test report to generate curve, and then print for analysis and archiving
Automatic short circuit protection to avoid equipment damage
Imported material fixture, high accuracy, easy and smart to use
Can simultaneously test two groups (6), three (9), four (12) concrete block one time
With the latest smart-in-one vacuum saturated machine
Test channels: 6 channels, 9 channels, 12 channels optional
Work environment: 0 centigrade ~ 50centigrade Humidity =< 75%
Measurement error: <1%
6.System configuration
Concrete Resist Chloride Ion penetration tester
Specimen fixture
concrete vacuum saturated machine(optional)
7.Test steps:
Cut the concrete Standard sample into 50mm thickness test block
The sample will be saturated with water in the intelligent concrete vacuum saturated machine
After the vacuum saturated, the specimens will be installed on the fixture
Inject the concrete solution (NaCl (negative), NaOH (positive) )into the specimen fixed poles
Connect the concrete Flux detector power and the positive and negative of specimen fixture
Connect the test host, test software control, the test results will be obtained after 6 hours
Data is automatically saved and can be printed.
Company Details
Bronze Gleitlager
,
Bronze Sleeve Bushings
and
Graphite Plugged Bushings
from Quality China Factory
Business Type:
Manufacturer,Exporter,Seller
Year Established:
2008
Ecer Certification:
Verified Supplier
Since 2008, Chongqing GOLD has developed 70 kinds of fire testing instruments and served many industries such as building materials, aviation, rails, IMO, wire & cable, safety protection, and so on.
Quality and Certification:We always adhere to a strict quality control system, and all products ... Since 2008, Chongqing GOLD has developed 70 kinds of fire testing instruments and served many industries such as building materials, aviation, rails, IMO, wire & cable, safety protection, and so on.
Quality and Certification:We always adhere to a strict quality control system, and all products ...