China factories

Chat Now Send Email
China factory - Xi'an Kacise Optronics Co.,Ltd.

Xi'an Kacise Optronics Co.,Ltd.

  • China ,Shaanxi
  • Verified Supplier
  1. Home
  2. Products
  3. About Us
  4. Contact Us

Leave a Message

we will call you back quickly!

Submit Requirement
China The KTG112-W Trace Oxygen Analyzer Is Applied To The On-line Analysis Of
China The KTG112-W Trace Oxygen Analyzer Is Applied To The On-line Analysis Of

  1. China The KTG112-W Trace Oxygen Analyzer Is Applied To The On-line Analysis Of

The KTG112-W Trace Oxygen Analyzer Is Applied To The On-line Analysis Of

  1. MOQ: 1PCS
  2. Price:
  3. Get Latest Price
Payment Terms L/C, D/A, D/P, T/T, Western Union, MoneyGram
Supply Ability 1000 Piece/Pieces per Week negotiable
Delivery Time 5-8 work days
Packaging Details each unit has individual box and all boxes are packed in standard packages or customers requests available
Measurement Range (0.0 - 10/100/1000) × 10⁻⁶;
Maximum Allowable Error ±2% FS (0.0 - 10) × 10⁻⁶; ±1% FS (10 - 1000) × 10⁻⁶;
Weight Approximately 1.5Kg.
Repeatability ±1% FS;
Sample Gas Flow Rate (400 ± 20) mL/min;
Response Time τ90 ≤ 60s;
Brand Name KACISE
Model Number KTG112-W
Certification CE
Place of Origin CHINA

View Detail Information

Inquiry by Email Get Latest Price
Chat online Now Ask for best deal
  1. Product Details
  2. Company Details

Product Specification

Payment Terms L/C, D/A, D/P, T/T, Western Union, MoneyGram Supply Ability 1000 Piece/Pieces per Week negotiable
Delivery Time 5-8 work days Packaging Details each unit has individual box and all boxes are packed in standard packages or customers requests available
Measurement Range (0.0 - 10/100/1000) × 10⁻⁶; Maximum Allowable Error ±2% FS (0.0 - 10) × 10⁻⁶; ±1% FS (10 - 1000) × 10⁻⁶;
Weight Approximately 1.5Kg. Repeatability ±1% FS;
Sample Gas Flow Rate (400 ± 20) mL/min; Response Time τ90 ≤ 60s;
Brand Name KACISE Model Number KTG112-W
Certification CE Place of Origin CHINA
High Light On-line Trace Oxygen AnalyzerAir Separation Trace Oxygen AnalyzerNitrogen Production Trace Oxygen Analyzer

The KTG112-W trace oxygen analyzer is applied to the on-line analysis of nitrogen production from air separation

 

Introduction

The KTG112-W trace oxygen analyzer is a new type of intelligent industrial portable analyzer developed by combining the electrochemical detection principle, imported high-performance electrochemical sensors, and new microcomputer technology.

Application Areas

Portable and rapid analysis of trace oxygen in nitrogen production from air separation, chemical processes, protective gases in high-temperature sintering furnaces for magnetic materials, protective gases in the electronics industry, as well as in the glass and building materials industries and various mixed gases.

 

Instrument Features

  • Selects imported fuel cell sensors, featuring long service life, high accuracy, and fast response, and is not interfered with by other gas components.
  • Has automatic temperature compensation to eliminate the influence of ambient temperature.
  • Optional components such as precision filters and air pumps can be added.
  • Uses a 128×64 dot-matrix OLED liquid crystal display, which can simultaneously show parameters such as oxygen content, date, time, and the internal temperature of the instrument.
  • Has a timed automatic storage function, allowing users to view stored data at any time.
  • Alarm points can be set arbitrarily within the full measurement range.
  • Can be powered by both AC and DC, offering flexible use. The battery can work continuously for about 10 hours after a full charge, and it will automatically alarm when the battery is under - voltage.

 

Technical Parameters

Measurement Range: (0.0 - 10/100/1000) × 10⁻⁶;

Maximum Allowable Error: ±2% FS (0.0 - 10) × 10⁻⁶; ±1% FS (10 - 1000) × 10⁻⁶;

Stability: Zero drift: ±1% FS/7d; Span drift: ±1% FS/7d;

Repeatability: ±1% FS;

Sample Gas Flow Rate: (400 ± 20) mL/min;

Response Time: τ90 ≤ 60s;

Resolution: 0.1 × 10⁻⁶;

Sample Gas Pressure: 0.05 MPa ≤ Inlet pressure ≤ 0.25 MPa;

Operating Environment: Temperature: -15°C - +45°C; Humidity: ≤90% RH;

Operating Power Supply: Rechargeable battery built - in the instrument;

Charging Power Supply: (220 ± 22) VAC, (50 ± 5) Hz. The charger has a built - in charging protection function;

Overall Dimensions: 255mm (Width) × 103mm (Height) × 265mm (Depth)

Weight: Approximately 1.5KG.

 

Company Details

Bronze Gleitlager

,

Bronze Sleeve Bushings

 and 

Graphite Plugged Bushings

 from Quality China Factory
  • Business Type:

    Manufacturer

  • Year Established:

    2008

  • Total Annual:

    8000000-10000000

  • Employee Number:

    100~150

  • Ecer Certification:

    Verified Supplier

Xi'an Kacise Optronics Co.,Ltd. is a professional manufacturer of  measuring instruments , located in China xi’an City. Since 2008, Xi'an Kacise focused on the development and production of pressure, level and flow measuring instruments .   We have been specializing... Xi'an Kacise Optronics Co.,Ltd. is a professional manufacturer of  measuring instruments , located in China xi’an City. Since 2008, Xi'an Kacise focused on the development and production of pressure, level and flow measuring instruments .   We have been specializing...

+ Read More

Get in touch with us

  • Reach Us
  • Xi'an Kacise Optronics Co.,Ltd.
  • i City, No11, TangYan South road, Yanta District, Xi'an,Shaanxi,China.
  • https://www.fluid-meter.com/

Leave a Message, we will call you back quickly!

Email

Check your email

Phone Number

Check your phone number

Requirement Details

Your message must be between 20-3,000 characters!

Submit Requirement