| Payment Terms | T/T | Model Number | RLK600 Series |
| Certification | CE; ISO | Brand Name | Gatherstar (Leading) |
| Place of Origin | China |
RLK600 digital infrared radiometer adopts mature photoelectric detection and digital processing technology, designed for field range tests. It has 4 models with different sensitivity and detection ranges, covering 2–5μm / 2–14μm spectrum. All optical and signal modules are integrated inside the probe, connected to computer via single RS-422 cable. Standard CCD image aiming and 5-inch LCD are configured, with switchable field of view and replaceable filters. Matched Windows software completes measurement control, data processing and calibration. All equipment passes Class 1 national defense metrology station calibration, widely used for engine, infrared decoy and thermal source characteristic testing.
Connect or disconnect the device (probe head), run device self-check and set device parameters
Start data collection automatically or manually, set collection duration
Auto name, save and back up data
Display collected data and curves in real time (synchronous image recording)
Open three types of data files: original backup data, formal test data and calibrated data
Data preprocessing (truncation, filtering, data recovery), automatic background removal and automatic calibration
Multiple graphic display modes (original data / background-removed data / calibrated data; zoom in / zoom out)
Import, modify calibration parameter files and re-calibrate; export data in TXT or Excel format
Support multiple radiation units: radiance, radiant intensity, irradiance
Calculate product calibration parameters automatically and form calibration files according to infrared radiation theoretical formulas and collected standard blackbody data.
![]()
| Model | RLK601 | RLK602 | RLK602D | RLK603 |
|---|---|---|---|---|
| Product Feature | Low sensitivity | Medium sensitivity | Medium sensitivity, dual-channel | High sensitivity |
| Application | Short-distance target detection (less than 5m) | Medium & long-distance target detection (5m-50m) | Long-distance target detection (more than 50m) | |
| Detector | Uncooled MCT | 2-4 stage electrically cooled PV-MCT | 2-4 stage electrically cooled PV-MCT Liquid nitrogen cooled or Stirling cooled InSb or HgCdTe | |
| Spectral Response | 2-14μm | 2-5μm or 2-14um | 2-5μm or 2-14μm | |
| NEI (Noise Equivalent Irradiance) | 5E-3W/cm*cm | 5E-8W/cm*cm | 5E-12W/cm*cm | |
| Dynamic Range | 1E+5 | 1E+5 | 1E+5 | |
| Linearity | <5% | <3% | <2% | |
| Stability | <5% | <3% | <2% | |
| Optical Input Diameter | 38mm(1.0”) | 38mm(1.65”) | 152mm(6”) | |
| Optical System | None | Transmissive infrared optical system | Reflective infrared optical system | |
| Measuring Field of View | 5°\10°\20°\30° (Electric diaphragm, select via software) | 0.75°\1.5°\2.25°\ 3.0° (Electric diaphragm, select via software) | 2.5mrad \ 5°\10° (Electric diaphragm, select via software) | |
| Filter | 1 filter holder (manual replacement) | |||
| Aiming Mode | CCD image aiming (5° FOV) with 5-inch LCD screen | |||
| Data Rate | 1Hz (high precision mode), 10Hz, 20Hz, 50Hz (fast measurement mode) (Signal modulation frequency: 1kHz) | |||
| Data Interface | RS-422 (Max transmission distance: 200 meters) | |||
| Operating Voltage | 220VAC / 24VDC | |||
| Operating Temperature | -40℃ ~ +50℃ | |||
| Dimension | RLK601/RLK602/RLK602D: 279(L) × 202(H) × 274(D) RLK603: 368(L) × 202(H) × 274(D) |
|||
| Weight | 8.0Kg (RLK601/RLK602/RLK602D) / 10Kg (RLK603) | |||
| Supporting Software | RLK600Manager (V4.0), special software for digital infrared radiometer Three functions: measurement control, data processing, product calibration | |||
| Optional Accessories | Tripod, laptop (standard / ruggedized), filters: 3.0-5.0μm, 8-14μm or customer specified; IRIG-B (DC) or IRIG-B (AC) time code input | |||
Note: Specifications are subject to change without prior notice, refer to official product manual.
Complete Transport Case
![]()
![]()
![]()
Q1: How many models are available for different testing distances?
A: Four grades: RLK601 (short-distance<5m), RLK602/602D (medium-long 5–50m), RLK603 (ultra-long>50m).
Q2: What is the maximum data sampling frequency?
A: Up to 50Hz fast sampling mode for dynamic radiation curve recording.
Q3: Can test data be exported for secondary analysis?
A: Yes, software supports TXT and Excel format data export with multiple radiation unit conversion.
Q4: Does the equipment support outdoor field measurement?
A: Equipped with complete portable aluminum case, wide -40℃~+50℃ operating temperature, tripod compatible.
Q5: Are customized optical filters available?
A: Standard 3.0–5μm & 8–14μm filters supplied, custom wavelength filters can be ordered as accessories.
Company Details
Business Type:
Manufacturer
Year Established:
2016
Total Annual:
10 million-50 million
Employee Number:
30~40
Ecer Certification:
Verified Supplier
Xi’an Gatherstar Optoelectronic System Co.,Ltd (Leading) is a state-certified high-tech manufacturer & integrated optoelectronic solution provider rooted in Xi’an, China, a city with world-class optical research resources. We integrate independent R&D, production, system integrat... Xi’an Gatherstar Optoelectronic System Co.,Ltd (Leading) is a state-certified high-tech manufacturer & integrated optoelectronic solution provider rooted in Xi’an, China, a city with world-class optical research resources. We integrate independent R&D, production, system integrat...
Get in touch with us
Leave a Message, we will call you back quickly!