| Payment Terms | L/C,T/T,D/P |
| Supply Ability | 100/Year |
| Delivery Time | 1-4 weeks |
| Packaging Details | Plywood case |
| Model NO. | YC99T-3C |
| Meter Position | 3 |
| Accuracy | 0.05 |
| Brand Name | GENY |
| Model Number | YC99T-3C |
| Certification | ISO, CE |
| Place of Origin | CHINA |
View Detail Information
Explore similar products
Portable Meter Test System For AC Voltage Current Power Phase And Frequency
Portable Meter Test System with High Accuracy Current and Voltage Source
Portable Meter Test System For Three Phase Energy Meters With Automatic
0.05 Accuracy Portable Meter Test Equipment Three Phase Test Meter Calibration
Product Specification
| Payment Terms | L/C,T/T,D/P | Supply Ability | 100/Year |
| Delivery Time | 1-4 weeks | Packaging Details | Plywood case |
| Model NO. | YC99T-3C | Meter Position | 3 |
| Accuracy | 0.05 | Brand Name | GENY |
| Model Number | YC99T-3C | Certification | ISO, CE |
| Place of Origin | CHINA | ||
| High Light | Portable meter test system with calibration ,Multifunction meter tester YC99T-3C ,Single-phase three-phase calibration meter | ||
The GENY Portable Meter Test System is a compact, high-precision solution for on-site and laboratory energy meter calibration. It is widely used by utility companies, power plants, metrology institutes, industrial enterprises, and energy meter manufacturers. The system supports testing of single-phase and three-phase energy meters, including active, reactive, composite, and multi-function smart meters.
Tests single-phase & three-phase mechanical/electronic meters
Automatic or manual testing: basic error, starting test, creeping test, standard deviation
Influence tests: voltage, frequency, harmonic, reverse phase sequence, unbalanced voltage
Supports DL/T645 communication protocol for multifunction smart meters
Multiple communication ports: RS232, RS485, Ethernet, USB, WiFi, I/O
Outputs 2–63 harmonics
8-inch touchscreen, fast and user-friendly operation
Tests up to 3 meters simultaneously
| AC Voltage Output | |
| Item | Specification |
| Voltage Range | 57.7V, 100V, 220V, 380V, 660V |
| Adjustment Range | 0–120% RG |
| Adjustment Resolution | 0.01%RG / 0.1%RG / 1%RG / 10%RG |
| Stability | 0.01%/2 min (0.05 cl), 0.02%/2 min (0.1 cl) |
| Distortion | ≤ 0.2% (Non-capacitive load) |
| Max Output Load | 20 VA / phase |
| Measurement Accuracy | 0.05%RG (0.05 cl), 0.1%RG (0.1 cl) |
| AC Current Output | |
| Item | Specification |
| Current Range | 0.05A, 0.2A, 1A, 5A, 16.67A, 50A |
| Adjustment Range | 0–120% RG |
| Adjustment Resolution | 0.01%RG / 0.1%RG / 1%RG / 10%RG |
| Stability | 0.01%/2 min (0.05 cl), 0.02%/2 min (0.1 cl) |
| Distortion | ≤ 0.2% (Non-capacitive load) |
| Max Output Load | 50 VA (50A range) |
| Measurement Accuracy | 0.05%RG, 0.1%RG |
| Power Output | |
| Item | Specification |
| Stability | 0.01%/2 min (0.05 cl), 0.02%/2 min (0.1 cl) |
| Active/Reactive Power Accuracy | 0.05%RG |
| Phase Output | |
| Item | Specification |
| Phase Adjusting Range | 0° – 359.99° |
| Adjustment Steps | 10°, 1°, 0.1°, 0.01° |
| Resolution | 0.01° |
| Measurement Accuracy | 0.05° |
| Power Factor Output | |
| Item | Specification |
| Adjustment Range | -1.0 ~ 0 ~ +1.0 |
| Measuring Resolution | 0.0001 |
| Measurement Accuracy | 0.0005 |
| Frequency Output | |
| Item | Specification |
| Adjustment Range | 40 – 70 Hz |
| Adjustment Resolution | 1 Hz / 0.1 Hz / 0.01 Hz / 0.001 Hz |
| Resolution | 0.001 Hz |
| Accuracy | 0.002 Hz |
| Harmonic Setting | |
| Item | Specification |
| Harmonic Order | 2nd–63rd |
| Harmonic Amplitude | 0–40% |
| Harmonic Angle | 0°–359° |
| Harmonic Setting Error | 2nd–31st: 0.1% ; 32nd–63rd: 0.2% |
| Energy Error Measurement | |
| Item | Specification |
| Active Energy Error Limit | 0.05%RD (30V–660V, 0.02A–60A, PF 0.5) |
| Reactive Energy Error Limit | 0.1%RD (30V–660V, 0.02A–60A, PF 0.5) |
| Standard Crystal | |
| Item | Specification |
| Frequency Accuracy | 0.5 ppm (factory @ 25°C) |
| Frequency Stability | 0.1 ppm (-20°C ~ +70°C) |
| Energy Pulse Input | |
| Item | Specification |
| Pulse Channels | 3 active + 3 reactive |
| General Parameters | |
| Item | Specification |
| Power Supply | AC220V ±15% or AC110V ±15% |
| Frequency | 50–60 Hz |
| Consumption | 60–1000 VA |
| Operating Environment | 20°C ~ 30°C, RH ≤ 85% |
| Storage Environment | -20°C ~ +50°C |
| Size | 600 mm (L) × 400 mm (W) × 266 mm (H) |
| Weight | 30 kg |
![]()
Company Details
Business Type:
Manufacturer,Exporter,Seller
Year Established:
1987
Total Annual:
2,000,000-10,000,000
Employee Number:
50~100
Ecer Certification:
Verified Supplier
Guangzhou GENY Electric Co. Ltd. is a leading manufacturer of energy meter test systems, whose predecessor is Guangzhou Yangcheng Science & Technology Enterprises Co., Ltd. which was established in 1987. We have been dedicated to the energy meter testing industry for over 30 years, and continuou... Guangzhou GENY Electric Co. Ltd. is a leading manufacturer of energy meter test systems, whose predecessor is Guangzhou Yangcheng Science & Technology Enterprises Co., Ltd. which was established in 1987. We have been dedicated to the energy meter testing industry for over 30 years, and continuou...
Get in touch with us
Leave a Message, we will call you back quickly!