Payment Terms | L/C, D/A, D/P, T/T, Western Union, MoneyGram |
Delivery Time | 3-5work days |
Operating Principle | Partial pressure electrochemical |
Output | 9 mV to 13 mV in 210 mBar O2 |
Measurement Range | 0 mBar to 1500 mBar O2 |
Response Time (T90) | <15 s (air to 100% O2) |
Baseline Offset | <200 μV |
Linearity | Linear 0% to 100% O2 |
Model Number | MOX4 AA829-M20 MOX-4 |
View Detail Information
Explore similar products
2FO AA625-180 Air Oxygen Sensor Combustion Efficiency For Fixed Gas Detector
AO2 PTB-18.10 AA428-210 Gas Oxygen Sensor Partial Pressure Electrochemical O2
MOX4 MOX-4 Oxygen Gas Sensor Medical Equipment White ABS AA829-M20 Oxygen
MAX250E R125P03-002 Electrochemical Dissolved Oxygen Sensor For Breathing
Product Specification
Payment Terms | L/C, D/A, D/P, T/T, Western Union, MoneyGram | Delivery Time | 3-5work days |
Operating Principle | Partial pressure electrochemical | Output | 9 mV to 13 mV in 210 mBar O2 |
Measurement Range | 0 mBar to 1500 mBar O2 | Response Time (T90) | <15 s (air to 100% O2) |
Baseline Offset | <200 μV | Linearity | Linear 0% to 100% O2 |
Model Number | MOX4 AA829-M20 MOX-4 | ||
High Light | Widely Used Oxygen Gas Sensor ,MOX4 Oxygen Gas Sensor |
MOX4 Oxygen Sensor
Part Number: AA829-M20
Document Purpose:
The purpose of this document is to present the performance specification of the MOX4 oxygen gas sensor.
This document should be used in conjunction with the Operating Principles (OP04) and the Product Safety Datasheet (PSDS 4).
The data provided in this document are valid at 20°C, 50% RH and 1013 mBar for 3 months from the date of sensor manufacture.
Output signal can drift below the lower limit over time. For guidance on the safe use of the sensor, please refer to the Operating Principles (OP04).
If the Sensor is Dropped:
If a sensor is dropped, then it should be placed in quarantine for 24 hours and a follow-up check made by a 2-point calibration.
Mechanical Installation:
When installing the sensor, it must only be screwed in hand-tight and a gas tight seal ensured. Spanners and similar mechanical aids may not be used, as excessive force may damage the sensor thread.
RFI/EMI Susceptibility:
MediceLs contain metal and may be susceptible to RFI or EMI. They are not suitable for use in MRI environments. For further information, please contact City Technology.
Housing Material | White ABS |
Weight | 42 g (nominal) |
Orientation | Any |
Typical Applications | Critical care anesthesia |
Operating Temperature Range | -20°C to +50°C |
Company Details
Business Type:
Manufacturer,Trading Company
Year Established:
2006
Total Annual:
900000-1000000
Employee Number:
50~80
Ecer Certification:
Verified Supplier
Yijiajie Electronics is a full range of silicon pressure sensors from the United States, FREESCALE/, GE, MEAS, SMI, and a full range of sensing and control products from the United States pressure, temperature and humidity, gas flow, current, Hall effect, etc., Japan HAMAMATSU flame, UV sensors, ph... Yijiajie Electronics is a full range of silicon pressure sensors from the United States, FREESCALE/, GE, MEAS, SMI, and a full range of sensing and control products from the United States pressure, temperature and humidity, gas flow, current, Hall effect, etc., Japan HAMAMATSU flame, UV sensors, ph...
Get in touch with us