Payment Terms | T/T, Western Union, Paypal |
Delivery Time | 3-5 working days |
Packaging Details | Standard Export Packing |
Brand | PHiliph Goldway |
Product Catgeory | Monitoring |
Product Type | Spare Parts |
Item Type | Board |
Item Name | ECG Board |
Model | GW4F4E01C |
Color | Green |
Size | Small |
Application | UT4000F |
Condition | Used - Good |
Brand Name | Goldway |
Model Number | UT4000F |
View Detail Information
Explore similar products
Patient Monitor Accessories Goldway UT4000 Monitor Blood Pressure Plate Great
Mindray PM7000 Patient Monitor Accessories Keyboard Silicone 7000-20-24395
Well-chosen Patient Monitor Mainboard Goldway G30 40 60 70 80 UT4000B UT6000
Patient Monitor Repair Parts GE Dash4000 Patient Monitor Flat Cable in ready
Product Specification
Payment Terms | T/T, Western Union, Paypal | Delivery Time | 3-5 working days |
Packaging Details | Standard Export Packing | Brand | PHiliph Goldway |
Product Catgeory | Monitoring | Product Type | Spare Parts |
Item Type | Board | Item Name | ECG Board |
Model | GW4F4E01C | Color | Green |
Size | Small | Application | UT4000F |
Condition | Used - Good | Brand Name | Goldway |
Model Number | UT4000F | ||
High Light | UT4000F Patient Monitor ECG Board ,GW4F4E01C Patient Monitor ECG Board |
Goldway UT4000F Monitor ECG Board
PHiliphh Goldway UT4000F Patient Monitor ECG Board GW4F4E01C Medical Spare Parts
► Brief Description
PHiliphh Goldway UT4000F Patient Monitor ECG Board GW4F4E01C
Goldway UT4000F Monitor ECG Board - This board is used, but is fully functional and in good condition. It can connect to the motherboard and other parts and receive instructions from the motherboard. When ECG monitoring is performed on the patient, this board will start to work accordingly.
This board will be carefully inspected before shipment to ensure it meets your usage requirements. All products of our company have a 90-day warranty. If you have any quality problems with this Goldway UT4000F Monitor ECG Board after you receive it, you can contact us at any time for processing.
► Specification
Brand: | PHiliphh Goldway |
Product Category: | Monitoring |
Product Type: | Spare Parts |
Item Type: | Board |
Item Name: | ECG Board |
Application Site: | ECG |
Application Machine: | UT4000F |
Model: | GW4F4E01C |
Color: | Green |
Size: | Small |
Condition: | Used - Good |
Warranty: | 90 Days |
► More Details
Generally, before starting to test the patient, an ECG test will be performed. The steps are as follows:
1. First, you need to adjust the ECG simulator to a heart rate of 80bpm and an ECG gain of 1mV;
2. Connect one end of the ECG cable and lead wire cable to the ECG connector interface of the UT4000F patient monitor, and the other end to the simulator;
3. Observe the ECG channel to see if there is any murmur in ECG lead II. When the R wave appears, there will be a "beep" sound;
4. Check whether there is ECG waveform and noise interference in each ECG lead; adjust the ECG gain to 2mV and observe whether there is any change in the ECG waveform;
5. Unplug the RA from the simulator and observe whether the screen prompts "lead off";
6. Reconnect the RA lead and set the 1mV calibration, and observe whether the 1mV calibration square wave overlaps. The ECG test is completed.
► ECG Performance Specifications
Patient Safety Standard | IEC60601-1-1988 |
CMRR | ≥60dB (Common Mode Rejection Ratio) |
Measure Range | 15 ~ 300bpm ± 1bpm |
Heart Rate Averaging | 8 second average |
ST Segment Range | -0.8 ~ + 0.8mV |
Interface | AAMI 6-pin |
Lead Selection | I, II, III (3 lead mode) I, II, III, aVR, aVL, aVF, V (5 lead mode) |
Lead Fault Alarm | Audible, Visual |
Input | 5-lead ECG patient cable |
QRS Indicator | Audible and Visual Alert |
Waveform Storage | 6 minutes |
Sweep Speed | 12.5/25/50 mm/sec |
Gain Selection | 4mV, 2mV, 1mV, 0.5mV, 0.25mV,Auto |
Trends | 2 hours → 4 hours → 8 hours → 24 hours → 48 hours |
Patient Isolation | Breakdown voltage: 4000VAC, 50Hz, 60 seconds, Leakage current: <10A |
Frequency width | Monitoring mode: 0.5 ~ 40Hz (+0.4dB,-3.0dB) Surgery mode: 0.5 ~ 20Hz (+0.4dB, -3.0dB) (not calibration significant) |
Patient Drive Current | < 10A |
Enclosure Leakage Current | < 0.1mA |
Maximum T wave Rejection Capability | 1.2mV |
Heart Rate Alarm Response Time | < 7 seconds |
Aspect Ratio | 0.24 ~ 0.6 sec/mV |
Defibrillator Protected & ESIS Protected | Tested with 5kV |
Recovery Time Following Defibrillation | < 5 seconds |
► Detailed Photos
Thanks for your interest in PHiliphh Goldway UT4000F Patient Monitor ECG Board GW4F4E01C Medical Spare Parts, please feel free to contact me. Thank you for your time and wish you have a happy life.
Company Details
Business Type:
Manufacturer,Distributor/Wholesaler,Agent,Importer,Exporter,Trading Company,Seller
Year Established:
2015
Employee Number:
20~50
Ecer Certification:
Verified Supplier
China World Technology Medical Equipment Service Group is a company engaged in medical equipments/repair parts and accessories products trading company. The company was established in September 2005, has many famous enterprises at home and abroad (such as America USCO company) to establish a long-te... China World Technology Medical Equipment Service Group is a company engaged in medical equipments/repair parts and accessories products trading company. The company was established in September 2005, has many famous enterprises at home and abroad (such as America USCO company) to establish a long-te...
Get in touch with us
Leave a Message, we will call you back quickly!