Ecer asks for your consent to use your personal data to:
Personalised advertising and content, advertising and content measurement, audience research and services development
Store and/or access information on a device
Your personal data will be processed and information from your device (cookies, unique identifiers, and other device data) may be stored by, accessed by and shared with 135 TCF vendor(s) and 65 ad partner(s), or used specifically by this site or app.
Some vendors may process your personal data on the basis of legitimate interest, which you can object to by do not consent. Contact our platform customer service, you can also withdraw your consent.
{"title":"Patient Monitor Repair Parts GE Dash4000 Patient Monitor Flat Cable in ready","imgUrl":"https:\/\/img.chinax.com\/nimg\/4b\/1f\/2f84f1d9346c0940eb1ca85ae447-200x200-1\/patient_monitor_repair_parts_ge_dash4000_patient_monitor_flat_cable_in_ready.jpg","attrs":{"Brand Name":"GE","Model Number":"Dash 4000","Minimum Order Quantity":"1pcs","Payment Terms":"T\/T, Western Union, Paypal"}}
{"title":"5 Leads Patient Monitor Parts Pm 7000 Portable Medical Monitor","imgUrl":"https:\/\/img.chinax.com\/nimg\/68\/83\/b83f5eaf4525a1984896d75fb9a5-200x200-1\/5_leads_patient_monitor_parts_pm_7000_portable_medical_monitor.jpg","attrs":{"Brand Name":"Mindray","Model Number":"PM-7000","Minimum Order Quantity":"1pcs","Payment Terms":"T\/T, Western Union, Paypal"}}
{"title":"V100 Monitor Controller Board , PWB Patient Monitor Accessories","imgUrl":"https:\/\/img.chinax.com\/nimg\/00\/3a\/a592d46380dd88bf6c663a31c7d8-200x200-1\/v100_monitor_controller_board_2c_pwb_patient_monitor_accessories.jpg","attrs":{"Brand Name":"GE","Model Number":"PWB 2034402-001 REV B","Minimum Order Quantity":"1","Payment Terms":"T\/T, Western Union,Paypal"}}
Mindray IPM-9800 Monitor SPO2 Plate includes two models: 050-00046-00 and 051-000058-00. The 050-00046-00 focuses on signal processing, while the 051-000058-00 focuses on interface drivers. Together, they form a complete blood oxygen monitoring unit, the core hardware for accurate SpO₂ and PR measurements. Their materials and circuit design fully meet the stringent requirements of medical equipment for stability, interference immunity, and safety.
Product Specifications
Product name
SPO2 Plate
Model
Mindray IPM-9800
PN
050-00046-00 051-000058-00
Circuit board material
medical grade FR-4 epoxy glass cloth laminate
Shipping time
Different
Deposit
100% pay before shipping
Payment terms
Bank TT/ Paypal/Western Union
MOQ
1 Piece
Size
110mm*90mm*1.6mm
Weight
About 60g
Core Functions
Signal Acquisition and Preprocessing
Blood Oxygen Saturation Calculation
Probe Adaptation and Driver
Probe Status Diagnosis
Frequently Asked Questions
Q1: The device displays "SpO₂ signal weak." I've checked the probe and it's fine. Is the oximeter board faulty?
A1: Turn off the power and check if the power cord is securely plugged in. After powering on the device, enter maintenance mode and check the oximeter board self-test results. If it displays "Signal amplification circuit abnormality" (fault code E04), the oximeter board is faulty and should be replaced.
Q2: Can the two oximeter boards be replaced separately? Or must they be replaced together?
A2: They can be replaced separately, depending on the fault type: ① If "Probe cannot be recognized" or "Drive current abnormality" occurs, it's likely that the 051-000058-00 board is faulty and can be replaced alone. ② If "Oximeter calculation error is large" or "Weak signal cannot be calibrated" occurs, it's likely that the signal processing chip or filter circuit on the 050-00046-00 board is faulty and should be replaced alone. ③ After replacement, power on the device and perform "SpO₂ calibration" to ensure measurement accuracy meets requirements.
Q3: The blood oximeter board gets noticeably hot while operating (approximately 65°C to the touch). Is this normal?
A3: It's not normal. The normal operating temperature should be ≤60°C. Possible causes of abnormal heating: ① The heat sink may have fallen off or the thermal grease may have dried out. Disassemble the device to inspect the heat sink on the 050-00046-00 board and reapply thermal grease (0.5mm thick). ② The driver chip is overcurrenting. Measure the probe LED current to see if it exceeds 80mA.
China World Technology Medical Equipment Service Group is a company engaged in medical equipments/repair parts and accessories products trading company. The company was established in September 2005, has many famous enterprises at home and abroad (such as America USCO company) to establish a long-te... China World Technology Medical Equipment Service Group is a company engaged in medical equipments/repair parts and accessories products trading company. The company was established in September 2005, has many famous enterprises at home and abroad (such as America USCO company) to establish a long-te...