| Packaging Details | In Wooden Case Or Customized |
| Size Range | 0.2m*0.3m-2.4m*1.2m |
| Composition | Rmon-pbo-sio2 (B2O3) |
| Place Of Origin | China |
| Supply Ability | 2000 Pcs Per Month |
| Impact Resistance | High |
| Lead Equivalent | 44mmpb |
| Delivery Time | 5-8 Work Days |
| Shape | Square |
| Payment Terms | T/T |
| Transparency | High Visible Light Transmission |
| Certification | CE |
| Brand Name | Jovi |
| Model Number | 1-cf-15 |
View Detail Information
Explore similar products
6.71mmpb 25mm Radiation Resistant Protection Lead Glass Xray ECT Room
10mm Radiation Protection Lead Glass X Ray Proof Glass 1000mm X 2000mm Without
25mm 6.71mmpb Radiation Protection Lead Glass X Ray Shielding 200 X300mm
ZF3 20mm X Ray Protection Radiation Shielding Window Glasses 1000mm X1500mm
Product Specification
| Packaging Details | In Wooden Case Or Customized | Size Range | 0.2m*0.3m-2.4m*1.2m |
| Composition | Rmon-pbo-sio2 (B2O3) | Place Of Origin | China |
| Supply Ability | 2000 Pcs Per Month | Impact Resistance | High |
| Lead Equivalent | 44mmpb | Delivery Time | 5-8 Work Days |
| Shape | Square | Payment Terms | T/T |
| Transparency | High Visible Light Transmission | Certification | CE |
| Brand Name | Jovi | Model Number | 1-cf-15 |
| High Light | 44Mmpb Lead Glass ,Glass Radiation ,Professional Lead Glass Radiation | ||
Jovvi Lead Glass for X-ray, CT, SPECT, PET‑CT & Nuclear Medicine – Version 2 (Comparative Selection Guide)
1. Quality – Performance Across Photon Energy Ranges
Different modalities produce different photon energies, and Jovvi lead glass quality is defined by how consistently it performs across this spectrum.
Key quality metrics:
| Parameter | X‑ray (≤150 kVp) | SPECT (140 keV ⁹⁹ᵐTc) | PET‑CT (511 keV) |
|---|---|---|---|
| Typical Pb eq. needed | 1.8–2.5 mm | 2.5–3.0 mm | 4.0–5.0 mm* |
| Visible transmittance | 84–88% | 80–84% | 68–74% |
| Lead oxide content | 55–60% | 60–65% | 65–70% |
| Physical thickness | 8–14 mm | 14–18 mm | 22–28 mm |
| Browning resistance | Standard | Enhanced CeO₂ | Maximum CeO₂ |
*For 511 keV, Jovvi often recommends composite windows (see Section 2).
Jovji ensures attenuation uniformity – variation across a 1.5 m panel is ≤ ±0.08 mm Pb equivalent at any energy.
2. Application Scenarios – Room‑by‑Room Guidance
A. Conventional X‑ray & CT (Diagnostic Radiology)
Typical setup: Single monolithic lead glass panel mounted in a lead‑lined drywall.
Jovvi recommendation: 2.0 mm Pb (X‑ray) or 2.3 mm Pb (CT) – balance of cost, weight, and visibility.
Special note: For CT rooms with 140 kVp or higher, use 2.5 mm Pb.
B. SPECT (Gamma Camera) Rooms
Photon energy: Mostly 140 keV (⁹⁹ᵐTc), also 59.5 keV (²⁴¹Am) and 364 keV (¹³¹I).
Jovvi recommendation: 2.8 mm Pb equivalent (physical ~16 mm).
Installation note: SPECT rooms often have the control booth adjacent to the gamma camera, requiring a window size of 800*600 mm to 1200*800 mm. Lead glass alone is sufficient if the wall shielding matches.
C. PET‑CT (511 keV Annihilation Photons)
Challenge: 511 keV photons have a half‑value layer in lead of approximately 4 mm. A single 5 mm Pb lead glass panel (physical ~28 mm thick) is heavy (~110 kg/m²) and expensive.
Jovvi composite solution (better value):
Option 1: 2.0 mm Pb lead glass (front) + 12 mm lead acrylic (middle) + 2.0 mm Pb lead glass (rear) – total shielding ~5.5 mm Pb eq. at 511 keV.
Option 2: 3.0 mm Pb lead glass alone (physical 18 mm) + additional 2 mm lead sheet on wall – lower weight but requires precise installation.
Visual quality: Composite windows have slight haze due to lead acrylic; Jovvi offers anti‑reflection coating to improve clarity.
D. Nuclear Medicine Hot Lab (Radiopharmacy)
Application: Viewing port into a fume hood or Laminar Flow Cabinet where ⁹⁹ᵐTc or ¹⁸F‑FDG is prepared.
Jovvi recommendation: Small panel (300*300 mm to 500*400 mm) with 2.5–3.0 mm Pb eq. – higher not needed because handling time is short and syringe shielding is used.
Quality requirement: Extremely low bubble count (<0.2 bubbles >0.3 mm per 100 cm²) to avoid visual obstruction during fine motor tasks.
E. Patient Isolation Room (High‑dose Therapy, e.g., ¹³¹I > 200 mCi)
Application: Viewing window from nurse station into patient room.
Jovvi recommendation: 3.5–4.0 mm Pb eq. Monolithic lead glass (physical 20–24 mm). Additional lead shielding on door and walls is more critical.
3. Production Assurance – Energy‑Specific Testing
Jovvi does not use generic “X‑ray testing" for nuclear medicine products. Each panel intended for SPECT or PET‑CT receives:
Attenuation measurement at 140 keV (using ⁹⁹ᵐTc source or X‑ray tube with 140 kVp + filtering) – required for SPECT rooms.
Optional 511 keV measurement – using a ²²Na source (annihilation photons). Jovvi provides this at extra cost for PET‑CT projects.
Uniformity scan – a 2D grid of attenuation values across the entire panel, ensuring no local weak spots.
Certificates included:
Standard certificate for X‑ray (IEC 61331‑2 compliant)
Nuclear medicine supplement (140 keV and/or 511 keV test results)
Optical distortion certificate (critical for PET‑CT where technologists view fusion images)
Logistics for nuclear medicine:
Jovvi can drop‑ship directly to hospitals worldwide with customs‑cleared documentation (HS code 7007.29).
Lead time: 30 days for up to 3.0 mm Pb; 45 days for 4.0–5.0 mm Pb or composite windows.
Company Details
Year Established:
2010
Total Annual:
10000000-20000000
Employee Number:
100~500
Ecer Certification:
Verified Supplier
Jovvi(Beijing Hengtai Tech Co., Ltd) was established in 2010. The company adheres to the development concept of “integrity all over the world and conducts all continents”, adheres to the business policy of “strive to create greater value for customers”, and is committed... Jovvi(Beijing Hengtai Tech Co., Ltd) was established in 2010. The company adheres to the development concept of “integrity all over the world and conducts all continents”, adheres to the business policy of “strive to create greater value for customers”, and is committed...