| Packaging Details | Wooden (Carton: Portable) |
| Delivery Time | 10-15 working days |
| Payment Terms | T/T |
| Supply Ability | 500 units/month |
| Brand Name | Senova |
| Place of Origin | Shanghai, China |
| Model Number | HCPS-D280 |
| Certification | CE |
View Detail Information
Explore similar products
200L Horizontal Cylindrical Autoclave HCPS-D200
150L Horizontal Cylindrical Autoclave HCPS-D150
280L Horizontal Cylindrical Autoclave HCPS-C280
400L Horizontal Cylindrical Autoclave HCPS-B400 with Stainless Steel Structure
Product Specification
| Packaging Details | Wooden (Carton: Portable) | Delivery Time | 10-15 working days |
| Payment Terms | T/T | Supply Ability | 500 units/month |
| Brand Name | Senova | Place of Origin | Shanghai, China |
| Model Number | HCPS-D280 | Certification | CE |
| High Light | 280L horizontal autoclave with warranty ,cylindrical autoclave for sterilization ,HCPS-D280 horizontal autoclave | ||
280L Horizontal Cylindrical Autoclave HCPS-D280
Features:
Specifications
| Model | HCPS-D280 |
| Chamber volume | 280L |
| Chamber size | φ600*1000mm |
| Working pressure | 0.22 MPa |
| Working temperature | 134℃ |
| Adjustment of temperature | 40℃ - 134℃ |
| Sterilization time | 0-99 min |
| Drying time | 0-99 min |
| Heat average | ≤± 2℃ |
| Power | 12KW/380V 50Hz |
| Dimensions | 1520*910*1900 |
| Packing size | 1680*1080*2100 |
| G.W/N.W(Kg) | 570/462 KG |
![]()
Introduction about autoclave
The term "autoclave" is synonymous with reliable sterilization in professional settings. This device operates on the principle that moist heat under pressure is microbicidal. By creating an environment of saturated steam at temperatures like 121°C or 134°C, it achieves what dry heat cannot at equivalent temperatures: rapid and deep penetration. The sterilization cycle is methodical.
First, the removal of air is critical; any residual air creates cold spots and prevents steam contact. Second, the exposure phase delivers the lethal conditions. Third, exhaust returns the chamber to ambient pressure, often aided by a vacuum to dry contents. This process is vital for sterilizing heat-resistant surgical tools, dental equipment, and a wide range of laboratory supplies. Laboratories depend on it to prepare culture media and decontaminate waste. Its use even spans into industrial sectors for materials processing. The rationale for its universal adoption is clear: it offers a non-toxic, rapid, and highly effective means of achieving sterility assurance, capable of inactivating the most persistent biological agents, which is why it remains the benchmark against which other methods are measured.
Company Details
Business Type:
Manufacturer,Exporter
Year Established:
2010
Employee Number:
50~100
Ecer Certification:
Verified Supplier
.gtr-container-senova456 { font-family: Verdana, Helvetica, "Times New Roman", Arial, sans-serif; color: #333; padding: 16px; line-height: 1.6; max-width: 100%; box-sizing: border-box; } .gtr-container-senova456 p { font-size: 14px; margin-bottom: 1em; ... .gtr-container-senova456 { font-family: Verdana, Helvetica, "Times New Roman", Arial, sans-serif; color: #333; padding: 16px; line-height: 1.6; max-width: 100%; box-sizing: border-box; } .gtr-container-senova456 p { font-size: 14px; margin-bottom: 1em; ...