Payment Terms | T/T,Western Union |
Delivery Time | 2-4 working days |
Packaging Details | Neutral Package |
Category | ultrasonic system |
Description | maintenance of ultrasound machine |
Service | ultrasound repair technician |
Warranty | 60 days after fixed |
Support | technical |
Condition | run failure |
Brand Name | GE |
View Detail Information
Explore similar products
Preirus Troubleshoot Ultrasound Machine Repair Hitachi Aloka
S2000 Color Doppler Medical Machine Repair For Startup Failure
GE Clinic Ultrasound Machine Repair
EPIQ 7 Ultrasound Machine Repair
Product Specification
Payment Terms | T/T,Western Union | Delivery Time | 2-4 working days |
Packaging Details | Neutral Package | Category | ultrasonic system |
Description | maintenance of ultrasound machine | Service | ultrasound repair technician |
Warranty | 60 days after fixed | Support | technical |
Condition | run failure | Brand Name | GE |
High Light | GE Vivid7 Ultrasound Machine Maintenance ,Run Failure Ultrasound Machine Maintenance |
GE vivid7 maintenance case
Fault case
Turn On the power and press On/Standby. The ultrasonic diagnostic device cannot start normally.
Fault analysis
Because Vivid 7 hardware is composed of the front end and the back end, its fault phenomenon is analyzed according to the activation sequence of Vivid 7.
(1) judge the machine is the front part or the back part of the fault caused by the failure of the normal start;
(2) the machine back end of a separate test, check whether it can start normally.The back end of the machine is mainly composed of microstar dual-core motherboard, ATX power supply, AGP graphics card, hard disk, ECG card, ETHERNET card, PCVIC board and PC2IP2B board.
Troubleshooting
(1) Remove the machine back-end components from the host, open the back-end chassis, input 220V ac power supply to the back end;
(2) Find the motherboard's pvon-on jumper, short circuit 6/8 feet (as shown in the picture)
(3) When it is found that the back-end part still cannot start normally, it is necessary to continue to confirm whether it is the problem of the motherboard or ATX power supply;
(4) Find the mainboard power board wiring block, with a short wire connection short circuit 13/14 (black and green) connector, at this time the back end of the power switch unimpeded;
(5) Input ac 220V into the back end, test the power voltage of each foot according to the ATX power sign, and find that there is no 3.3v voltage in the foot 1/2/11. It is judged that there is a problem in the ATX power supply at the back end. Repair the 3.3v power supply and test it on the machine after repairing the power.
Ultrasound parts that we can repair:
Repair List of Featured Ultrasonic Board & Assy | |||
Brand | System | Description | Part Number |
GE | Logiq P5 | CPU mainboard | 5168431 |
GE | Logiq P6 | CPU mainboard | |
GE | Logiq E9/vivid E9 | GFI2 | 5161631 |
GE | Voluson E6/Voluson E8 | RSR | KTI301394-2/KTI196357 |
GE | Voluson E6/Voluson E8 | RFI/ RFI21b Board | KTI300614/KTI302197-6 |
GE | Voluson S6/S8/P8 | BF64 | 5396937-2 |
GE | Voluson S6/S8/P8 | RFS board | 5364098-2/5364098-3 |
GE | Voluson S6/S10/P8 | BF192 board | 5357234 |
Aloka | Alpha 5 | mainboard | EP493700DD |
Aloka | Alpha 6 | RX Beam Former | EP555501AA |
Aloka | Alpha 6 | CW board | EP555700 |
Aloka | Alpha 6 | CPU mainboard | EP537000/EP563000 |
Aloka | Alpha 6 | TX/RX | EP555300 |
Aloka | Alpha 6/Alpha 7 | DOM | EP550000AA |
Aloka | Alpha 10 | RX Beamformer board | EP531800CE |
Aloka | Alpha 10 | wave generator Board | EP495000HH/EP495000HJ/EP495000FF |
Hitachi | F37 | TX/RS | EP557400/EP557300 |
Hitachi | F37 | mainboard | EP575700BC/EP560800 |
Aloka | F75 | USP board | EP556600 |
Aloka | F75 | Cell Board | EP558800 |
Hitachi | HI VISION Avius | RDBF board | 7513907A/7513612 |
Hitachi | Arietta 60 | CELL | EU-9160C |
Hitachi | Arietta 70 | TX | EP572300AA |
Hitachi | Arietta 70 | CELL | |
Hitachi | Noblus | CELL | |
Hitachi | Noblus | CONT-B | |
Hitachi | Ascendus | CELL | 7352830A |
Siemens | X150/G40 | BE | 10131804 |
Siemens | X300 | BE 2.0/BE 3.0/BE 7.5 | 10131990 |
Siemens | X700 | BE | 10136465 |
Siemens | X700 | RC | 10136467 |
Siemens | X700 | TI | 11014292 |
Siemens | Antares | BE | 7472421 |
Siemens | Antares | CB | 7288504 |
Siemens | S2000 | RM300 | 10852163 |
Siemens | S2000 high version/S3000 | RM301 | 10854749 |
iU22/iE33 | EMB excelsior motherboard | 453561419431/453561382231/453561419431 | |
iU22/iE33 | UMB Unified Motherboard | 453561254492/453561360251 | |
HD15 | ACB acquisition control board | 453561197145/453561197144 | |
HD15 | AIM board | 4.53561E+11 | |
CX50 | control panel | 4.53562E+11 | |
CX50 | mainboard | 453561368033/453561622163/453561495832 | |
Affiniti 70 | ACQ acquisition control board | ||
EPIQ 5/EPIQ 7 | ACQ Acquisition Module board | 453561734844/453561704246 | |
Biosound | Mylab Series | CPU mainboard | 9501361000 |
Toshiba | Aplio 300/400/500 | mainboard | PM30-38696 |
Toshiba | Aplio 500 | RX | ywm2056*a |
Toshiba | Xario 200 | TX | PM30-32732-1 |
Toshiba | Xario 200 | RX | PM30-32733-1 |
Remarks | |||
1.The repair cycle is within 3-5 days. | |||
2.Warranty after fixed:60 days. | |||
3. Kindly send us the estimated main faults (better to have image) for our engineer to prejudge. | |||
4.Package price is available if repair 3 pcs or more at one time. | |||
5.Extra discount for VIP partners. | |||
6.You could have different part No.under same description. |
Company Details
Business Type:
Distributor/Wholesaler,Exporter,Trading Company,Seller
Year Established:
2014
Total Annual:
5,000,000-10,000,000
Employee Number:
50~60
Ecer Certification:
Verified Supplier
Rongtao Medical was established in 2013, we have been focusing on providing after-sales service for medical equipment for the last ten years, especially in ultrasound. We have been continuously improving our technical capabilities, breaking through technical problems, innovating service models to ex... Rongtao Medical was established in 2013, we have been focusing on providing after-sales service for medical equipment for the last ten years, especially in ultrasound. We have been continuously improving our technical capabilities, breaking through technical problems, innovating service models to ex...
Get in touch with us
Leave a Message, we will call you back quickly!