| Payment Terms | L/C, D/A, D/P, T/T, Western Union, MoneyGram |
| Supply Ability | 10000 T/Month/Project |
| Delivery Time | 10 Days |
| Packaging Details | In: Bottle and Strip; Out: Box |
| Packaging | reagents bottles, reagents strip |
| Storage | 2~8℃ |
| CV | <=3% |
| Classification | Tumor Marker |
| Test throughput | 120+ tests/hour |
| Calibration | 6 points |
| Brand Name | Zecen |
| Model Number | CLIA, POClia |
| Certification | ISO 13485, CE, FSC, |
| Place of Origin | China, Taizhou |
View Detail Information
Explore similar products
Thyroid TSH Thyroid Stimulating Hormone For Automatic Immunoassay Analyzer In
Hypertension Test 100 Test Kit(Renin) for Automatic immunoassay analyzer In
Serum/Plasma/Whole Blood Test Estradiol Reagent kit E2 for Fertility Hormone In
Solution Reagent Kit PROG Fertility Hormone Progesterone Immunoassay Reagents
Product Specification
| Payment Terms | L/C, D/A, D/P, T/T, Western Union, MoneyGram | Supply Ability | 10000 T/Month/Project |
| Delivery Time | 10 Days | Packaging Details | In: Bottle and Strip; Out: Box |
| Packaging | reagents bottles, reagents strip | Storage | 2~8℃ |
| CV | <=3% | Classification | Tumor Marker |
| Test throughput | 120+ tests/hour | Calibration | 6 points |
| Brand Name | Zecen | Model Number | CLIA, POClia |
| Certification | ISO 13485, CE, FSC, | Place of Origin | China, Taizhou |
Tumor Marker CA50 Carbohydrate Antigen 50 for Fully Automatic Immunoassay Analyzer IVD Reagents 100T/box,24T/box
INTENDED USE
The kit has been designed for the quantitative determination of Cancer Antigen 50 (CA50) in human serum.
The method can be used for samples over the range of 1.0~140.0 U/mL.
The test has to be performed on the CIA fully auto analyzer (Including CIA 600, CIA 1200, CIA 1800, CIA2800).
PERFORMANCE CHARACTERISTICS
1) Precision
Intra-assay coefficient of variation was evaluated on testing Control 1 and Control 2 10 times each of 3 different batches in the same assay, then calculate the average and get the coefficient of variation.
| Intra-assay precision | |||
| Control | Mean(U/mL) | SD(U/mL) | CV% |
| Level 1 | 20.40 | 0.44 | 2.17% |
| Level 2 | 73.51 | 1.67 | 2.28% |
Inter-assay coefficient of variation was evaluated on three batches of kit. Repeatedly measured same level of control serum 10 times for each batch, calculating the coefficient of variation.
| Inter-assay precision | |||
| Control Level 1 | Mean(U/mL) 19.99 | SD(U/mL) 1.05 | CV% 5.24% |
| Level 2 | 78.86 | 5.12 | 6.49% |
2) Analytical Sensitivity
Test the CA50 zero standard 20 times Calculate the mean value (M)and standard deviation(SD) of the RLU and get the value of M+2SD. Then correspond it into the simple equation got by regression fitting from Calibrator A and Calibrator B.The sensitivity is typically less than 1.0 U/mL.
3) Specificity
The specificity of the CA50 assay system was assessed by measuring the apparent response of the assay to various potentially cross reactive analytes.
| Compound | Concentration | Cross reactivity |
| CEA CA19-9 | 150 ng/mL 400 U/mL | 0.12 U/mL 0.41 U/mL |
4) Accuracy (Fortified Recovery)
Use 10 μL CA50 sample A in concentration of 100U/mL(permitted deviation ±20%) add to 90 μL sample B which could be serum or other appropriate substrate. Test the diluted sample 3 times. Calculate the average concentration. The accuracy should be within 85%-115%.
| Sample | Mean Measuring | Fortified Recovery |
| 10A+90B | 21.74 U/mL | 91.0% |
5) Linearity
Dilute CA50 calibrator F by the ratio of 1/8, 1/16, 1/32, 1/64, 1/128. measuring all points’ RLU 3 times and calculate the mean value, and then do linear fitting by least square method. The absolute linear correlation coefficient(r) should be bigger than 0.9900.
| Calibrator Point | Concentration U/mL | Absolute linear correlation coefficient (r) |
| F | 139.62 | |
| 1/8F | 16.49 |
R= 0.9999
|
| 1/16F | 9.80 | |
| 1/32F | 5.33 | |
| 1/64F | 2.93 | |
| 1/128F | 1.12 |
![]()
Company Details
Business Type:
Manufacturer
Year Established:
2011
Employee Number:
300~500
Ecer Certification:
Active Member
Zecen Biotech CO., LTD., founded in 2011, is a leading Chinese diagnostics manufacturer specializing in in-vitro diagnostics devices and reagents. Zecen focus on the Chemiluminescent immunoassay (CLIA) field. With all products CE marked, our market presence currently spreads from across China to oth... Zecen Biotech CO., LTD., founded in 2011, is a leading Chinese diagnostics manufacturer specializing in in-vitro diagnostics devices and reagents. Zecen focus on the Chemiluminescent immunoassay (CLIA) field. With all products CE marked, our market presence currently spreads from across China to oth...
Get in touch with us
Leave a Message, we will call you back quickly!