| Water quality output | Type II or Type I ultrapure water |
| Compliance standards | FDA, USP, ISO 9001 |
| Sanitization method | Hot Water |
| Sanitization frequency | Daily or as required |
| Product name | Hot Water Sanitizable EDI Module |
| Operating temperature | Up to 85 °C |
| Flow rate | Varies by model, typically 1-10 GPM |
| Warranty period | 1 year standard |
| Module dimensions | Varies by model, typically compact size |
| Connection type | Standard sanitary fittings |
| Membrane material | High-grade ion exchange membranes |
| Maximum pressure | 100 psi |
| Electrical consumption | Low, depends on model |
| Application | Pharmaceutical, laboratory, and semiconductor industries |
| Module type | EDI (Electrodeionization) |
| Brand Name | Aquacells |
| Model Number | EIMD-50HP |
| Certification | FDA,EU,CE,RoHS ,REACH |
| Place of Origin | China |
View Detail Information
Explore similar products
Product Specification
| Water quality output | Type II or Type I ultrapure water | Compliance standards | FDA, USP, ISO 9001 |
| Sanitization method | Hot Water | Sanitization frequency | Daily or as required |
| Product name | Hot Water Sanitizable EDI Module | Operating temperature | Up to 85 °C |
| Flow rate | Varies by model, typically 1-10 GPM | Warranty period | 1 year standard |
| Module dimensions | Varies by model, typically compact size | Connection type | Standard sanitary fittings |
| Membrane material | High-grade ion exchange membranes | Maximum pressure | 100 psi |
| Electrical consumption | Low, depends on model | Application | Pharmaceutical, laboratory, and semiconductor industries |
| Module type | EDI (Electrodeionization) | Brand Name | Aquacells |
| Model Number | EIMD-50HP | Certification | FDA,EU,CE,RoHS ,REACH |
| Place of Origin | China | ||
| High Light | High Temperature Resistant EDI Module ,EDI Module 90-95% Recovery | ||
The Aquacells EIMD-HP module is specifically engineered for hot water sanitization applications and is widely utilized in Pharmaceutical, biotechnology, and cosmetics industries. Constructed from high-temperature resistant materials to accommodate instantaneous heating and cooling requirements, this module delivers a constant flowrate of high-purity water without the need for chemical regeneration.
| Parameter | Specification |
|---|---|
| FCE (including CO2 and Silica) | < 45 µS/cm |
| Temperature | 5-45 ℃ (40-113 ℉) |
| Feed pressure | 1.4-7 bar (20-100 psi) |
| Total Chlorine (as CI2) | <0.05 ppm |
| Fe | <0.01 ppm |
| Mn | <0.01 ppm |
| S2- | <0.01 ppm |
| pH | 4-11 |
| Total Hardness (as CaCO3) | <1.0 ppm |
| Dissolved Organics (TOC as C) | <0.5 ppm |
| Silica (SiO2) | <1.0 ppm |
| Parameter | Performance |
|---|---|
| Typical Recovery | 90-95% |
| Maximum Feed Pressure | 7 bar (100 psi) |
| Maximum Feed Temperature | 45 ℃ (113 ℉) |
| Sanitization Temperature at 30 psi (2.0 bar) | 85 ℃ (185 ℉) |
| DC Voltage* | 0-600 V |
| DC Amperage | 0-10 A |
| Product Conductivity | <0.1 µS/cm |
| Silica (SiO2) Removal | 90-99% depending on feed water |